Optical brighteners respond to UV, so a lamp with the wrong UV content gives wrong brightness and whiteness on brightened papers even when non-fluorescent tiles read perfectly. The operator measures a fluorescent reference standard and moves a UV-adjusting filter, or lets the software tune it, until the reading matches the certified value for illuminant C or D65. It is separate from the ordinary white calibration and must be redone whenever the lamp or filter changes.
New lamp went in Monday and nobody redid the UV calibration, so every OBA sheet this week reads two points hot.
A non-fluorescent tile cannot reveal a UV calibration problem; only a fluorescent standard will.