Imported uncertainty

nounStandard termMetrology

The uncertainty carried into a measurement from the calibration certificate of the reference standard used.

The uncertainty component taken from the calibration of the reference standard or instrument used to make a measurement, typically read off its certificate as an expanded uncertainty and divided by the stated coverage factor. It is the link that carries uncertainty down the traceability chain, so no lab can claim a smaller uncertainty than the one it imports. Drift of the standard since that calibration belongs on a separate budget line.

Heard on the job

Imported uncertainty from the NMI cert is already 0.5 ppm, so a 0.3 ppm CMC claim can't be right.

Also heard as

  • reference standard uncertainty
  • uncertainty of the standard